Quick answer
ResMed’s AirSense 10 and 11 guides say to wash tubing weekly in warm water with a mild detergent or dishwashing liquid, then rinse it well. Let it dry fully, out of direct sunlight and heat. Never put it in a dishwasher or washing machine.
Your CPAP tubing carries air from your machine to your mask. This guide covers what ResMed’s AirSense 10 and AirSense 11 user guides say about washing and drying it, including ClimateLineAir heated tubing. If you’re not sure which machine you have, see How to find your machine and mask model name.
What your user guide says
Both AirSense user guides put the air tubing in the weekly cleaning steps. ResMed’s U.S. cleaning page agrees: the mask cushion and water tub are daily jobs, and the tubing is a weekly one. Here are the steps, in ResMed’s order.
- Unplug your machine.
- Hold the cuff at the machine end and gently pull the tubing away from the machine.
- Hold the cuff and the mask swivel, then gently pull them apart.
- Wash the tubing, inside and outside, in warm water with a mild dishwashing liquid.
- Rinse it well in clean water.
- Let it dry out of direct sunlight and heat.
- Once it’s dry, connect it firmly to the air outlet on the back of the machine. Then connect the other end to your mask.
The AirSense 11 and ClimateLineAir 11 guides say not to wash tubing in water hotter than 131°F (55°C). ResMed’s U.S. cleaning page suggests warm, drinking-quality water, about 86°F (30°C).
Tip: Pull on the cuff, not the tube. ResMed’s ClimateLineAir 11 guide says holding or pulling the tubing itself may damage it.
The AirSense 11 guide offers a vinegar-and-water option for the water tub and outlet connector. For the air tubing, it lists only warm water and a mild dishwashing liquid. For the tub, see How to clean your CPAP water chamber.
Heated tubing (ClimateLineAir 11)
ClimateLineAir 11 heated tubing is also washed once a week. Its user guide gives the same steps: warm water, household dishwashing liquid, a thorough rinse, then drying out of direct sunlight and heat.
ResMed adds two notes. The end with the electrical connector fits only the air outlet on the machine, never your mask. And the AirSense 11 guide says all parts should be dry before you plug the machine back in.
ResMed lists ClimateLineAir 11 for the AirSense 11, and ClimateLineAir (no “11”) for the AirSense 10. Using heated tubing on an AirSense 10? Follow the user guide that came with your tubing.
What not to do
- No dishwasher or washing machine. Both AirSense guides say not to wash the tubing in either one.
- No harsh cleaners. ResMed’s guides say not to use bleach, chlorine, aromatic-based solutions, moisturizing or antibacterial soaps, or scented oils. They say these may cause damage and shorten the life of the parts.
- No alcohol or strong scents. ResMed’s U.S. cleaning page also lists alcohol and strong-smelling products, like citrus soaps. It says leftover vapors can be harmful to breathe in.
Ozone and UV cleaners
The U.S. Food and Drug Administration (FDA) says there are no FDA-cleared or approved devices for cleaning, disinfecting or sanitizing CPAP machines. It also says UV light and ozone can break down CPAP accessories like hoses and masks. And the FDA says it has no evidence from makers that ozone or UV light can clean the inside of CPAP hoses.
The FDA notes that it has authorized one add-on device for certain masks and hoses. It says an add-on device does not replace the maker’s cleaning steps.
ResMed’s guides add that ozone may damage the machine, and its limited warranty won’t cover that damage. ResMed’s U.S. cleaning page says its user guides call for mild detergent and warm water only.
How to dry your tubing
Both AirSense guides say to let the tubing dry out of direct sunlight and heat. ResMed’s U.S. cleaning page suggests laying it flat on a clean towel. The FDA says CPAP hoses can also be hung up to air dry.
Make sure it’s dry before you reconnect it. ResMed’s AirSense 10 and ClimateLineAir 11 guides say to put the tubing back once it is dry.
When to replace it instead
Washing can’t fix damage. Both AirSense guides say to check your tubing and replace it if it has any holes, tears or cracks.
The AirSense 11 and ClimateLineAir 11 guides go a step further. If a part shows visible wear, such as cracking, discoloration or tears, they say to throw it out and replace it.
ResMed’s guides put this check after cleaning and before you reconnect the tubing. For how often tubing is usually replaced, see How often should you replace CPAP supplies? To find the right new tube, see Standard, SlimLine or heated tubing: which fits?
When to call your supplier
Call the company that supplied your equipment if:
- You’re not sure which tubing fits your machine.
- You need new tubing and insurance pays for your supplies. Your supplier can tell you what your plan needs.
- Water drops keep showing up in your tubing or mask. Don’t change your machine’s settings on your own. Ask your supplier.
- Your machine is damaged, makes unusual sounds, or doesn’t work the way it usually does. ResMed’s AirSense 11 guide says to stop using it. Call your supplier.
Parts for this guide
Check the label on your current part, or your user guide, before you order.
-
Tubing & hoses ResMed
ResMed ClimateLineAir 11 Heated Tubing
$61.99Member price $56.99 How membership works Add to cart -
Tubing & hoses ResMed
ResMed Standard CPAP Tubing, 6.5 ft
$42.99Member price $37.99 How membership works Add to cart -
Cleaning & care
CPAP Hose Cleaning Hanger
$28.99Member price $23.99 How membership works Add to cart
Sources
- ResMed. AirSense 11 User Guide (398291, Americas edition), “Cleaning and caring for the device” and “Troubleshooting.” https://document.resmed.com/documents/products/machine/airsense-11/user-guide/398291_airsense-11_user-guide_amer_multi.pdf
- ResMed. AirSense 10 User Guide (3781122, Americas edition), “Caring for your device.” https://document.resmed.com/documents/products/machine/airsense-series/user-guide/3781122_airsense-10_user-guide_amer_multi.pdf
- ResMed. ClimateLineAir 11 User Guide (398178, Americas edition), “Setting up” and “Cleaning and caring for your tubing.” https://document.resmed.com/documents/products/accessory/climatelineair-11/user-guide/climatelineair-11_user-guide_amer.pdf
- ResMed. Cleaning CPAP equipment: Everything you need to know (U.S.). https://www.resmed.com/en-us/sleep-health/resources/cleaning-cpap-equipment/
- U.S. Food and Drug Administration. Do You Need a Device That Claims to Clean a CPAP Machine? (content current 08/26/2024). https://www.fda.gov/consumers/consumer-updates/do-you-need-device-claims-clean-cpap-machine
- ResMed. CPAP humidification and tubing (U.S.). https://www.resmed.com/en-us/products/cpap/replacement-parts/components-accessories/humidification-tubing/